50m³ Walk In Temperature Humidity Environment Stability Test Chamber

Brand Name:Sinuo
Certification:Calibration Certificate (Cost Additional)
Model Number:SN886-50m³
Minimum Order Quantity:1
Delivery Time:45 days
Payment Terms:T/T
Contact Now

Add to Cart

Verified Supplier
Location: Guangzhou Guangdong China
Address: Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone, Shibi Street, Panyu District, Guangzhou, China
Supplier`s last login times: within 28 hours
Product Details Company Profile
Product Details

50m3 Walk In Temperature Humidity Environment Stability Test Chamber


Complied Standards & Clauses


IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”


IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle) , IDT”


IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”


IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”


Equipment Overview


Samples and Restrictions: Electrical, electronic, mechanical and large parts, semi-finished products, finished products, etc.


Applications: It is used for high and low temperature and damp heat tests on electricians, electronics, machinery, and large parts, semi-finished products, and finished products. Process quality and reliability testing of products, parts, etc. under simulated temperature and humidity conditions.


Test Principle: The heating, humidification and refrigeration system is used to control and change the ambient temperature and humidity in the chamber, simulating the ambient temperature / humidity and alternating procedures required by the testing.


Structure: Adopts closed chamber structure.


Technical Parameters


Control & OperationPLC and touch screen intelligent control system
Chamber Size

50m3: WDH=5×2×5m

Can be customized

Temperature Range (Optional)

Maximum test temperature: 80 ° C, 120 ° C;

Minimum test temperature: -60 ° C, -40 ° C, -20 ° C, 0 ° C

Temperature Deviation≦±2.0℃
Temperature Fluctuation±0.5℃
Temperature Uniformity±2℃
Humidity Range30%〜98%RH
Humidity Deviation±5.0%RH
Heating Rate≤2.5°C/min (nonlinear, no load)
Cooling RateAverage 1.0 ° C / min
Temperature and Humidity Range Icon
System Main Composition
Chamber BodyThe inner chamber is made of SUS304# stainless steel plate, rounded shape, smooth, easy for clean
The outer chamber is made of A3 cold-rolled steel plate electrostatic double-sided spray or SUS304 stainless steel
Lead hole: inner and outer chambers are equipped with a threading hole with a rubber stopper
Insulation material: polyurethane foam
Observation window: double-layer explosion-proof tempered glass door
Heating SystemHeating method: electric heating
High quality nickel-chromium wire electric heater, contactless control (SSR)
Humidification SystemHumidification method: external humidification structure
Cooling SystemGerman Bock semi-tight compressor or German BITZER semi-compact compressor
Non-fluorine environmentally friendly refrigerant R404A R23, in line with environmental regulations, safe and non-toxic.
Shell and tube condenser (water cooled).
High-performance fin type automatic load capacity adjustment, long-term use of low temperature and high humidity conditions without frosting.
Control SystemTouch screen: Weinview programmable temperature and humidity controller (7 inch 16-bit true color touch screen)
Temperature control: PID self-tuning, using Class A PT100 platinum resistance sensor

Function: fixed value test and program test

Fixed value test: When the test environment requires only a single temperature and humidity, selecting this mode to set the required temperature, humidity and running time.

Program test: When the test environment requires multiple temperatures and humidity, select this mode to set the temperature, humidity and running time of each program. User can set the program running sequence. The program group is up to 100 groups.

Data StorageWith temperature and humidity curve interface, with standard R232 interface or U disk communication interface, the data can be downloaded, and the data changes and processes can be recorded.
Safety ProtectionPower-off memory function, phase sequence protection, leakage protection, over-temperature protection, compressor overload protection, water shortage alarm, abnormal protection
Safety Protection DeviceHeater air-burning protection switch, heater over current circuit breaker, circulating fan over current overload protection
Compressor high voltage protection switch, compressor overheat protection switch, compressor over current protection switch
Under-reverse phase circuit breaker, no-fuse switch, zero-crossing gate fluid power controller
Alarm IndicationWhen the above protection occurs, the equipment stops running, an audible and visual alarm is issued, and the fault location is displayed on the meter.

China 50m³ Walk In Temperature Humidity Environment Stability Test Chamber supplier