Portable 8 Inch Multi Parameter Patient Monitor Vehicle

Brand Name:SUCE
Certification:ISO13485
Model Number:PM-9000
Minimum Order Quantity:1
Delivery Time:5-8 days
Payment Terms:D/A, D/P, T/T, Western Union, MoneyGram,Paypal,Payoneer
Contact Now

Add to Cart

Active Member
Location: Hong kong China
Address: ROOM 702,7/F, SPA CENTRE,NO. 53-55 LOCKHART ROAD,WAN CHAI, HONG KONG
Supplier`s last login times: within 1 hours
Product Details Company Profile
Product Details

Portable 8 inch vehicle multi - parameter monitor

Details:


1. New high-precision module, new digital oxygen technology, more comprehensive measurement, more accurate.


2. High brightness high resolution 8 inch color TFT liquid crystal display, exquisite design, weight only 1.3kg.


3. With ST segment automatic detection, arrhythmia analysis, drug concentration drop table analysis.


4. Full-parameter trend chart data, ecg waveform storage and playback, power failure saving function.


5. Rotating shuttle operation mode, Chinese and English logo button, easy to learn, easy to operate, simple and intuitive.


6. Coexistence of standard, large font, full lead and trend chart is convenient for multi-angle remote viewing.


7. Built-in removable lithium battery, AC/DC dual-use, lightweight, energy saving and environmental protection.


8. Suitable for adults, children and newborns; It is suitable for bedside monitoring, ICU/CCU, operating room, emergency center, home monitoring, etc.


9. Optional four vehicle-mounted devices: suitable for all kinds of transport vehicle-mounted mobile environment.


Standard configuration: 1 PM-9000 8-inch vehicle monitor host, one built-in rechargeable lithium battery, one standard five-lead heart lead connection cable, one standard adult blood pressure cuff, one blood oxygen probe, one temperature probe, one power plug cable, one operating manual, one set of three certificates.


Technical parameters:


ECG ECG


Lead selection: full lead, ⅰ, ⅱ, ⅲ, aVF, aVR, aVL, V1 -- 6 chest lead display


Gain selection: ×1, ×2, ×0.25, ×0.5 four grades


Frequency: 0.5 Hz - 100 Hz


Common mode signal rejection: greater than 70dB


Scanning speed: 12.5mm/s, 25mm/s, 50mm/s three standard files


Accuracy of heart rate: ±5 beats/min


Calibration signal: 1mV (peak value, accuracy ±3%)


ST segment detection: Measurement range (-2.0mV -- +2.0mV)


Use standard: comply with ANSI/AMI EC13-2002 standard


Response time: 1S


Alarm mode: sound and light alarm


Preset range of heart rate alarm :(alarm over preset value)


The upper limit can be adjusted from 1 to 255 times/min


The lower limit can be adjusted from 0 to 254 times/min


NIBP blood pressure


Technique: oscillographic method


Measurement method: Reverse measurement data is more accurate


Measurement unit: mmHg/Kpa


Overvoltage protection: double safety protection


Measurement time: less than 45S


Cuff pressure range: 0-250mmHg


Accuracy of blood pressure measurement: mean error and standard deviation values specified in ANSI/AAMI SP10-2002


Initial charge of cuff: 180mmHg (120mmHg for children)


Cuff follow-up inflation: original systolic +30mmHg


Blood pressure measurement range and error:


Systolic pressure 60 mmHg to 250 mmHg


The average pressure is 45 mmHg to 235 mmHg


Diastolic pressure 40 mmHg to 220 mmHg


Blood pressure measurement error: ± 1.1kpa (±8mmHg) or ±10% whichever is greater


Response time: 1S


Alarm mode: sound and light alarm


Alarm prediction range and error :(alarm over preset value)


The upper systolic blood pressure ranges from 1mmHg to 255 mmHg


The lower limit ranges from 0 mmHg to 254 mmHg


The upper diastolic blood pressure ranges from 1mmHg to 255 mmHg


The lower limit ranges from 0 mmHg to 254 mmHg


Alarm error: ±1 mmHg


Standard: comply with ANSI/AAMI SP10-2002 standard


SPO2 oxygen saturation


Scope: 0-100%


Accuracy: 1%


Resolution: 1%


Pulse rate: 20-250 beats/min


Response time: 1S


Alarm mode: sound and light alarm


Alarm range and error :(alarm over preset value)


The upper limit is 1% to 100%


The lower limit ranges from 0% to 99%


TEMP temperature


Technology: high sensitivity thermistor probe


Temperature measurement range: 20℃-- 45℃


Resolution: 0.1%


Temperature accuracy: ±1℃


Response time: 1S


Alarm mode: sound and light alarm


Preset alarm range and error :(alarm over preset value)


The upper limit is adjustable between 20.1℃ and 45.5℃


The lower limit can be adjusted between 20℃ and 45.4℃


Alarm error: ±1℃


Breathe RESP


Respiratory rate range: 0-60 breaths/min


Accuracy: + / - 1 RPM


Response time: 1S


Gain selection: ×1, ×2, ×0.5, ×0.25 four grades


Alarm mode: sound and light alarm


Preset alarm range and error :(alarm over preset value)


The upper limit is adjustable between 10-100 beats/min


The upper limit is adjustable between 0 and 99 beats/min


Alarm error: ±1rpm


System


Display: 8 inch TFT true color LCD screen


Battery (optional) : LeadAcid fully sealed and maintenance-free battery


Safety standard: LEC601-1 (INTERNATIONAL GB9706.1)

China Portable 8 Inch Multi Parameter Patient Monitor Vehicle supplier

Portable 8 Inch Multi Parameter Patient Monitor Vehicle