Oxygen flow range | 1-10L/Min 1-15L/Min |
Pressure reducing range | 2.0~3.0kg |
Relief valve Pressure range | 3.5kg±0.5kg |
Mode | pneumatically powered and electrically controlled system |
Working Mode | Closed; Semi-Closed; Semi-Open |
Circuit | Breathing circuit integrated standards |
Tube | 4Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min |
Trolley | Fitted with 4 nos anti-static rubber castors; two of which are
lockable for braking and easy maneuverability with foot operated brake provisions |
Drawer unit | ONE drawer totally extractable |
Gas requirement: | Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa. | |
safety valve | <12.5 kPa | |
Respiratory Rate | 1~99bpm | |
Flow meter | O2:0.1~10L/Min, N2O:0.1~10L/Min | |
oxygen concentration in the mixed gas N2O/O2 | > 21% | |
Oxygen Flush | 25~75 L/min | |
modes of ventilation | A/C , IPPV, SIPPV, IMV, SIMV, MANUAL | |
I/E Ratio | 4:1 ~ 1:10, Has the inverse ratio ventilation | |
Tidal Volume | 0~1500ML | |
ptr inspiratory triggering pressure | -0.1kPa~1.0kPa | |
Inspiratory Plateau | 0~1s | |
Maximum safety pressure | ≤ 12.5 kPa | |
Pressure limit range | 0 ~ 6.0 kPa | |
Airway pressure alarm: Audible and visual and with yellow and red
color indicating | Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa | |
±0.2 kPa | ||
Tidal volume alarm: Audible and visual and with yellow and red
color indicating | upper alarm: 50 to 2000ml, lower alarm: 0~1800ml | |
Power Supply Alarm | Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s | |
The airway pressure continues to be higher than 15 hPa ±1 hPa for
15s±1s, then the machine will raise an audible alarm, the pressure
will be displayed in red and the continuous high pressure red alarm
wording is displayed on the screen of the anesthetic respirator. |