

Add to Cart
Introduction
The Flow Analyzer Go8, referred to hereafter as the Analyzer, is a
general-purpose gas flow analyzer. The Analyzer measures
unidirectional gas flow, airway pressure and high pressure. The
Analyzer can also measure tidal volume, inspiratory peak pressure,
positive
end-expiratory pressure, etc.
This manual provides a description of the Analyzer and its
applications. lt should be used as a guide when using the
instrument to analyze ventilators and test lungs.
Key Features
Unidirectional flow rate, airway pressure, low pressure, high
pressure and oxygen concentration measurement
Flow, pressure real-time waveform display
Touch buttons control
RS232 serial port and can be used for flow calibration
Intuitive interface, easy to operate
Easy to carry
Unpacking and inspection
Use the following checklist when unpacking the Analyzer. There are
no special unpacking instructions, but be careful not to damage the
instrument when unpacking it.
Check the shipping carton is in good condition. lf the shipping
carton is damaged, carefully continue unpacking the instrument and
note any dents and scratches on the Analyzer. Save the damaged
shipping carton and packing material for the carrier inspection.
lf there is no shipping damage, continue removing the Analyzer from
the shipping case. Save the box and packing materials.
Perform a visual inspection to ensure that the Analyzer is intact
and if there is any physical damage, such as bent or broken parts,
dents, or scratches. If the Analyzer has been damaged, call the
theSiriusmed Service Center immediately.
After unpacking the Analyzer, check to see if all of the standard
accessories listed under the packing list have been included with
the analyzer. if anything is missing, contact the Siriusmed
ServiceCenter.
Operating Environmental Conditions | |
Temperature Range | 10 - 40 °c |
Ambient Humidity | 0- 80 %non-condensing |
Barometric Pressure | 520-800 mmHg |
Storage Environmental Conditions | |
Temperature Range | -25 to 50°℃ |
Humidity | 0 -95 % non-condensing |
Power Requirements | |
lnput Voltage Range | 12V DC |
Power Consumption | 36 VA |
Flow Port | |
Flow Range | 0 -200 LPM |
Accuracy | 3% of reading or 0.1 LPM |
Resolution | 0.01 LPM flow <5 LPM;0.1 LPM flow >5 LPM |
Notes
Flow accuracy is specified for dry air or oxygen.
Do not use with humidified gases.
Flow accuracy is obtained by allowing the Analyzer to fully warm up.
Airway Pressure | |
Pressure Range | 0 - 100 cmH2O |
Accuracy | 3 % of reading or 0.5 cmH2O |
Resolution | 0.1 cmH2O |
High Pressure | |
Operating Pressure Range | 0 - 690 kPa |
Accuracy | 2 % of reading or 5 kPa |
Resolution | 0.1 kPa |
Barometric Pressure | |
Operating Pressure Range | 120 - 860 mmHg |
Accuracy | 2 % of reading or 5 mmHg |
Resolution | 1 mmHg |
Gas Temperature | |
Temperature Range | 0 - 70 °C |
Accuracy | 2 %FS |
Resolution | 0.1 °C |
Oxygen Concentration(Optional) | |
Range | 0 - 100 % |
Accuracy | 2 %FS |
Resolution | 0.1 °C |