HbA1c Test Analyzer LD-100 POCT Testing Analyzer Fully Automatic
Hematology Detection Analyzer
Intended Use
LD-100 (POCT) One-Step HbA1c Testingf Analyzer is intended for the
determination of hemoglobin A1c(HbA1c) content in human whole
blood. It is a semi-automatic instrument used to analyze glycated
hemoglobin which can be used to detect of HbA1c in medical units at
all levels in daliy life. LD-100 analyzer uses the fluorescent
immunoturbidimetry, he percentage of glycosylated hemoglobin in
total hemoglobin can be determined.
Specifications
Model | LD-100 |
Device category: | Fixed |
Power connection | Detachable power supply cord |
Measurement Time | Less than 3 minutes |
Sample Volume | 4ul |
Sample Type | Whole blood |
Data storage | 7000 |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
Analytical wave-length | 510nm and 580nm. |
Certification | NGSP, IFCC |
PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an
integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the
buffer and mixed to release the hemoglobin from the red blood
cells.
- The total concentration of hemoglobin depends on the initial
reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and
glycosylated hemoglobin is determined by detecting the decrease of
fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can
be determined.
Feature
- One-Step Testing, Automatic Detection
- Protable and convenient
- Fast test speed and High Accuracy
- Battery Integrated
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen without extra instrument
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge
Performance Parameters
- Accuracy: Is tested with the reference material as samples, and the relative
deviation of the determination results of the analyzer should be
within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample
coefficient of variation (CV) of the test results is not more than
3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation
coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the
relative deviation of the test results is not more than ±3%.
- Functions of the analyzer:The function of report output in English; Abnormal alarm function.
Certificated
Company Profile
Labnovation Technologies, Inc. is an original manufacturer
established in Shenzhen, China since 2001. We are dedicated in
developing and manufacturing IVD instruments and reagents including
HbA1c analyzer and reagents, rapid test reader & reagents,
hematology reagents, clinical chemistry reagents, electrolyte
reagents, wash solution and urine test reagents.
Labnovation products are developed in house, tested extensively in
clinics and passed strict quality controls. All of our products are
CE marked and approved by SFDA. In the past 15 years, our products
have been exported to 110 different countries and used in 8000
hospitals worldwide. We provide both ready-to-use final products
and bulk OEM reagents, including ready-to-use reagents,
concentrated, and powder.
With an investment of 10% of our annual sales into R&D, we
release new products every years. We win more and more domestic and international cooperation by
company reputation and products quality while we actively take part
in fairs like MEDICA,AACC, ARBA HEALTH and so on through which we
are well-known by the companies in this field.